Time of analysis is an important factor for investigations and a determining factor for fast processes. What is the fast and ultra-fast Raman imaging? Everything has to be compared and the terminology doesn’t matter.

Minimal time of acquisition of Raman spectrum for each pixel of an image is basically limited by CCD detector operation speed (about: 760 µs – 1 ms) and makes in practice about 5 ms. In Confotec® NR500 system we apply two modes for Raman imaging. Alongside with CCD detection mode we apply the mode in which PMT is a receiver and the scanning is realized with galvanic mirrors at a fixed sample. We call this mode Fast mapping, rather modestly, considering that the time of acquisition of each pixel of the image is only 3 µs! And the minimal time of 1001 pixels x 1001 pixels Raman image acquisition is only 3 s!

Some capabilities of Fast mapping mode have been shown at its application to the investigation of Granite Gneiss India sample (the sample has been supplied by Brno University of Technology). Fast mapping mode allows imaging both at relatively strong signals and maximal signals which are several orders weaker, if compared with the Raman signal from silicon. Two dry objectives have been applied for the measurements: 100х and 20х.

As an example the confocal (1AU) megapixel (1002001 pixels) images of Granite Gneiss India sample with total measurement time of 3 seconds are shown in Fig.1. The images in Fig.1 have been obtained with the objective 100х by scanning of maximum sample size e for the given objective (150 μm х 150 μm) at scanning step of 150 nm.

Fig.1a. Sample surface image in the reflected laser light on the wavelength of 488 nm
Fig.1a. Sample surface image in the reflected laser light on the wavelength of 488 nm.
Fig.1b. Raman image of anatase (titanium dioxide) distribution
Fig.1b. Raman image of anatase (titanium dioxide) distribution.
Fig.1c. Summarized image of anatase distribution relative to the image of the sample surface in the reflected light
Fig.1c. Summarized image of anatase distribution relative to the image of the sample surface in the reflected light.
Fig.1d. Summarized axonometric image
Fig.1d. Summarized axonometric image.

Parameters of images in Fig.1

  • System: Confotec® NR500
  • Mode: Fast mapping
  • Sample: Granite Gneiss India
  • Scanning field: 150 μm х 150 μm
  • Number of pixels: 1002001
  • Scanning step: 150 nm
  • Time per 1 pixel: 3 µs
  • Registration time: 3 s
  • Objective: 100х dry

It is possible to choose only an informative part of the obtained image and realize the scanning of the appropriate part of the object with the same number of pixels. As a result, a more detailed image with higher spatial resolution will be obtained. In Fig.2a the examples of such megapixel (1002001 pixels) images scanned with the step of 43 nm are given. The informative part of the object can be scanned also in a high sensitivity mode. Thus the image with higher ratio signal/noise will be obtained. Examples of such Fast mapping application for imaging are shown in Fig.2b.

Original image for areas 2a and 2b

Parameters Fig.2a

  • System: Confotec® NR500
  • Mode: Fast mapping
  • Sample: Granite Gneiss India
  • Scanning field: 43 μm х 43 μm
  • Number of pixels: 1002001
  • Scanning step: 43 nm
  • Time per 1 pixel: 3 mcs
  • Registration time: 3 seconds
  • Objective: 100х dry

Parameters Fig.2b

  • System: Confotec® NR500
  • Mode: Fast mapping
  • Sample: Granite Gneiss India
  • Scanning field: 43 μm х 43 μm
  • Number of pixels: 63001
  • Scanning step: 172 nm
  • Time per 1 pixel: 48 mcs
  • Registration time: 3 seconds
  • Objective: 100х dry

Fig.2a

Fig.2a.1. Anatase distribution image of the area 2a
Fig.2a.1. Anatase distribution image of the area 2a.
Fig.2a.2. Laser image of area 2a
Fig.2a.2. Laser image of area 2a.
Fig.2a.3. Summed image of anatase distribution relative to the image of sample surface in reflected light
Fig.2a.3. Summed image of anatase distribution relative to the image of sample surface in reflected light – area 2a.
Fig.2a.4. Axonometry of 2a.3
Fig.2a.4. Axonometry of 2a.3.

Fig.2b

Fig.2b.1. Anatase distribution image of the area 2b
Fig.2b.1. Anatase distribution image of the area 2b.
Fig.2b.1. Anatase distribution image of the area 2b
Fig.2b.2. Laser image of area 2b.
Fig.2b.3. Summed image of anatase distribution relative to the image of sample surface in reflected light – area 2b
Fig.2b.3. Summed image of anatase distribution relative to the image of sample surface in reflected light – area 2b.
Fig.2b.4. Axonometry of 2b.3
Fig.2b.4. Axonometry of 2b.3.

Anatase spectrum in one of the points of the image (100х objective) is given in Fig.3. The Raman line used for imaging is schematically marked with a rectangle.

Fig.3. Anatase spectrum in one of the points of image 1 (100x objective)
Fig.3. Anatase spectrum in one of the points of image 1 (100x objective).

Photoluminescence from the studied material or from sample impurities is a problem interfering with the detection of Raman images. In Confotec® NR500 Raman confocal microscope a specially developed system of removing the photoluminescence is applied for the imaging in Fast mapping mode. Its application is successful regardless of photoluminescence appearance – in the form of separate bright spots distorting Raman image, in the form of photoluminescence background in the whole image or in the case of both photoluminescence appearance simultaneously.

In Fig.4a the anatase distribution obtained with the 20x objective in Fast mapping mode with distortions induced by the photoluminescence in the form of separate bright spots is shown. Separate photoluminescence spots are marked with circles for visual demonstration.

In Fig.4b a real anatase distribution is shown after the photoluminescence distortions have been removed. The image of the same sample area but in CCD detection mode (Mapping mode) after use of Baseline Correction function for the photoluminescence removing is shown for comparison in Fig.4c. Both images have no distortions induced by the photoluminescence. Anatase spectrum in one of the points of sample (20x objective) is shown in Fig.4d.

Figure 4

Demonstration of removing interfering photoluminescence in the form of separate spots.

Fig.4a. Fast mapping image distorted by the photoluminescence
Fig.4a. Fast mapping image distorted by the photoluminescence. Some photoluminescence spots are marked with circles for visual demonstration.
Fig.4b. Real anatase distribution without photoluminescence traces
Fig.4b. Real anatase distribution without photoluminescence traces (251 pixel x 251 pixel).
Fig.4c. Image of the same sample area obtained in CCD detection mode after use of Baseline correction function for the photoluminescence subtraction
Fig.4c. Image of the same sample area obtained in CCD detection mode after use of Baseline correction function for the photoluminescence subtraction.
Fig.4d. Anatase spectrum in one of the image pixels (20x objective)
Fig.4d. Anatase spectrum in one of the image pixels (20x objective).

Another example of photoluminescence subtraction that totally distorts Raman image and appear both in the form of separate bright spots and in the form of the photoluminescence background in the whole image is shown in Fig.5.

Figure 5

Example of photoluminescence subtraction that totally distorts the image.

Fig.5a. Initial image scanned with Fast mapping mode (101 pixel x 101 pixel ) with 20x objective lens
Fig.5a. Initial image of quartz distribution scanned with Fast mapping mode (101 pixel x 101 pixel ) with 20x objective lens.
Fig.5b. Fast mapping image (101 pixel x 101 pixel) of real quartz distribution with no photoluminescence impact
Fig.5b. Fast mapping image (101 pixel x 101 pixel) of real quartz distribution with no photoluminescence impact.
Fig.5c. Image of the same sample area scanned with resolution of 21 pixel x 21 pixel obtained with the CCD detection mode after subtraction of photoluminescence
Fig.5c. Image of the same sample area scanned with resolution of 21 pixel x 21 pixel obtained with the CCD detection mode after subtraction of photoluminescence.
Fig.5d. Quartz spectrum in one of the image pixels (20x objective lens)
Fig.5d. Quartz spectrum in one of the image pixels (20x objective lens).

Figures 6-8 show the examples of Fast mapping mode application for investigation of Granite Gneiss India with 20x objective.

Parameters Fig.6

  • System: Confotec® NR500
  • Mode: Fast mapping
  • Sample: Granite Gneiss India
  • Scanning field: 194 μm х 194 μm
  • Number of pixels: 10201
  • Scanning step: 1.94 nm
  • 1.94 nm: 4.4 mcs
  • Registration time: 45 s
  • Objective: 20х dry

Figure 6

Fig.6a. Image obtained at Raman shift ~240 1/cm
Fig.6a. Image obtained at Raman shift ~240 cm-1.
Fig.6b. Image of the same sample area at Raman shift ~1353 1/cm
Fig.6b. Image of the same sample area at Raman shift ~1353 cm-1.
Fig.6с. Spectrum measured in one of image pixels with Raman lines used for analysis
Fig.6с. Spectrum measured in one of image pixels with Raman lines used for analysis.

Figure 7

  • System: Confotec® NR500
  • Mode: Fast mapping
  • Sample: Granite Gneiss India
  • Scanning field: 194 х 194 μm
  • Number of pixels: 63001
  • Scanning step: 0.776 μm
  • Time per 1 pixel: 0.7 ms
  • Registration time: 45 s
  • Objective: 20х dry
Fig.7a. Image of quartz distribution
Fig.7a. Image of quartz distribution.
Fig.7b. Spectrum in one of the image pixels with the marked Raman line used for analysis
Fig.7b. Spectrum in one of the image pixels with the marked Raman line used for analysis.

Figure 8

  • System: Confotec® NR500
  • Mode: Fast mapping
  • Sample: Granite Gneiss India
  • Scanning field: 194 х 194 μm
  • Number of pixels: 63001
  • Scanning step: 0.776 μm
  • Time per 1 pixel: 48 µs
  • Registration time: 3 s
  • Objective: 20х dry
Fig.8a. Image of anatase distribution
Fig.8a. Image of anatase distribution.
Fig.8b. Summed image of anatase distribution in relation to the image of the sample surface in the reflected laser light
Fig.8b. Summed image of anatase distribution in relation to the image of sample surface in the reflected laser light.
Fig.8с. Spectrum in one of image pixels
Fig.8с. Spectrum in one of image pixels.

The advantages of using the Fast mapping mode:

  • High speed of confocal Raman image scanning – up to 3 µs/pixel.
  • High spatial resolution. Minimal scanning step – less than 21.5 nm.
  • Possibility of selection of resolution depending on size of the investigated area: 101 pixel x 101 pixel, 251 pixel х 251 pixel, 501 pixel х 501 pixel, 1001 pixel х 1001 pixel.
  • High sensitivity. A special mode for increasing the signal/noise ratio at measurement of weak Raman signals.
  • Combination of high speed, high sensitivity and high resolution allows the fast acquisition of reliable data on distribution of the chemical compound over a sample. When such high-speed and sensitive mode as our Fast mapping is not available the following technique is used for acquisition of chemical compound distribution over a sample: at first a large sample area is measured with a low resolution, and then the informative areas are scanned with higher resolution. However at low resolution a specific compound in a sample may not be found at all, especially if this chemical compound has a form of separate inclusions in a sample.
  • Raman images acquisition with photoluminescence subtraction.
  • Channel Reflection Fast Mapping for acquisition of the same high-speed images with high spatial resolution of the same investigated sample areas but in the reflected laser light. Besides the auxiliary information about a sample obtained in the reflected laser light this channel quickly finds the focused areas of the relief surface of a sample.
Author: Alexander Gvozdev, Leading research engineer
Publication date: September 4, 2013

You may also be interested in:

contact us