Compact IMAGING spectrograph SL100M

The SL100M device is a simple and compact lens spectrograph with astigmatism compensation, intended for operation in ultraviolet, visible and near-infrared spectral regions. SL100M is designed on a horizontal-symmetric configuration with two objectives, focal length of 100 mm and F/number of 1/3.3. Wavelength scanning is performed manually with a micrometer screw.

The distinctive features of the SL100M Imaging spectrograph manufactured by SOL instruments are: high optical throughput, plane focal field, manually-regulated entrance slit, compact size, ease of use and very light weight. Due to compensation of astigmatism and distortion, the SL100M device is ideal for application in multi-channel spectroscopy and for luminescent measurements.

Other spectrographs from SOL instruments:

Help & Support

contact us

Description of compact spectrograph SL100M

Small-sized Imaging spectrograph SL100M

Astigmatism-compensated Imaging lens spectrographs of SL100M series with focal length of 100 mm and F/number of 1/3.3 are designed on a horizontal-symmetric configuration with two objectives. Wavelength scanning is performed manually with a micrometer screw. Due to compensation of astigmatism and distortion, the SL100M devices are perfect for application in multi-channel spectroscopy.

The standard delivery set of the SL100M spectrographs includes a camera adapter with «C-mount» flange. Other flanges for CCD detectors connection can be supplied on a customer’s request.

Application areas of SL100M

  • Multi-channel spectroscopy
  • Luminescent measurements

Optical scheme of SL100M

Оптическая схема компактного спектрографа SL100M

Features of SL100M

  • High optical throughput and plane focal field
  • Manually-regulated entrance slit
  • Compact, light-weight and easy to operate

Specification of Imaging spectrographs of SL100M series

Optical system

Optical scheme:horizontal-symmetrical,
with astigmatism compensation
F/number:1/3.3
Focal length:99.3 mm
Objectives:achromatized, distortion-compensated
Flat field:28 х 8 mm (20 х 8 mm with C-mount adapter)
Diffraction grating size:40 x 40 x 6 mm
Stray light
(20 nm from 632.8 nm laser line):
9 х 10-4
  • Optical scheme: horizontal-symmetrical, with astigmatism compensation
  • F/number: 1/3.3
  • Focal length: 99.3 mm
  • Objectives: achromatized, distortion-compensated

  • Flat field: 28 х 8 mm (20 х 8 mm with C-mount adapter)
  • Diffraction grating size: 40 x 40 x 6 mm
  • Stray light (20 nm from 632.8 nm laser line): 9 х 10-4

Imaging

Horizontal magnification:1.2
Vertical magnification:1
Vertical spatial resolution:50 µm
  • Horizontal magnification: 1.2
  • Vertical magnification: 1
  • Vertical spatial resolution: 50 µm

Drive

Control:manual, with micrometer screw
Mechanism:sine
Reading accuracy (micrometer):1 nm
  • Control: manual, with micrometer screw
  • Mechanism: sine
  • Reading accuracy (micrometer): 1 nm

Optics

Device model:SL100MSL100M-UV
Diffraction grating:400 l/mm 500 nm400 l/mm 300 nm
Wavelengths range limited by the constructive parameters:360 – 1100 nm200 – 1100 nm
Wavelengths range with diffraction grating efficiency >40%:265 – 1000 nm200 – 600 nm
Reciprocal linear dispersion (at blaze wavelength):18.38 nm/mm19.21 nm/mm
Spectral resolution:< 0.35 nm< 0.37 nm
Repeatability:± 0.1 nm± 0.1 nm
Wavelength accuracy:± 1 nm± 1 nm

Model: SL100M

  • Diffraction grating: 400 l/mm 500 nm
  • Wavelengths range limited by the constructive parameters: 360 — 1100 nm
  • Wavelengths range with diffraction grating efficiency is greater than 40%: 265 — 1000 nm
  • Reciprocal linear dispersion (at blaze wavelength): 18.38 nm/mm
  • Spectral resolution: less than 0.35 nm
  • Repeatability: ±0.1 nm
  • Wavelength accuracy: ±1 nm

Model: SL100M-UV

  • Diffraction grating: 400 l/mm 300 nm
  • Wavelengths range limited by the constructive parameters: 200 — 1100 nm
  • Wavelengths range with diffraction grating efficiency is greater than 40%: 200 — 600 nm
  • Reciprocal linear dispersion (at blaze wavelength): 19.21 nm/mm
  • Spectral resolution: less than 0.37 nm
  • Repeatability: ±0.1 nm
  • Wavelength accuracy: ±1 nm

Spectral slits

Control:manual, with micrometer screw
Opening width:smoothly regulated
from 0 up to 2.0 mm
Parallelism:± 1 µm
Accuracy (slit width 1 mm):± 10 µm
Repeatability:± 1.5 µm
Reading accuracy (micrometer):2 µm
  • Control: manual, with micrometer screw
  • Opening width: smoothly regulated from 0 up to 2.0 mm
  • Parallelism: ±1 µm
  • Accuracy (slit width 1 mm): ±10 µm
  • Repeatability: ±1.5 µm
  • Reading accuracy (micrometer): 2 µm

Dimensions

Overall sizes (L x W x H):190 х 175 х 160 mm
Weight:2 kg
  • Overall sizes (L x W x H): 190 х 175 х 160 mm
  • Weight: 2 kg

Dimensional drawing of spectrograph SL100M

Dimensions of spectrograph SL100M