МС 122 / МС 124
Description
Specification
Accessories
Applications
Device modelМС 122МС 122ММС 122АМС 124МС 124-1
Operating spectral rangefrom 190 up to 1100 nmfrom 315 up to 1100 nmfrom 380 up to 1100 nm
Spectral slit width3 nm1,4 nmfrom 0,3 up to 3,0 nm
(step 0,1 nm)
3 nm1 nm
Resolution for toluene in hexane> 1,7> 1,9
Optical scheme typemonochromator with double dispersion, double-beam (split beam)
Light sourcesdeuterium and halogen lampshalogen lamp
Photodetectorssilicon photodiodes
Photometric rangeby optical density: from -0,3 up to 4 В; by transmittance: from 0,01 up to 200 %
Wavelength setting errornot more than ± 0,5 nmnot more than ± 0,3 nmnot more than ± 0,3 nmnot more than ± 0,5 nmnot more than ± 0,3 nm
Wavelength setting reproducibilitynot more than 0,2 nmnot more than 0,1 nmnot more than 0,1 nmnot more than 0,2 nmnot more than 0,1 nm
Wavelength calibrationautomaticautomaticautomatic;
by built-in filter
automaticautomatic;
by built-in filter
Margin of error measuring transmittance T± 0,3 %± 0,3 % (with a spectral slit width 1,5 nm)± 0,3 %
Standard deviation measuring transmission Tnot more than 0,05 %
(Т = from 1 up to 100 %)
not more than 0,1 %
(Т = from 1 up to 100 %)
not more than 0,1 %
(Т = from 1 up to 100 %, spectral slit width 1,5 nm)
not more than 0,1 %
(Т = from 1 up to 100 %)
Zero drift (500 nm)not more than ± 0,0005 В/hour
Interfering radiation levelnot more than 0,03 %
(220, 340, 400 nm)
not more than 0,02 % (220, 340, 400 nm)
not more than 0,5 % (200 nm)
not more than 0,03 %
(340, 400 nm)
not more than 0,02 %
(400 nm)
Cuvette compartmentinside dimensions 120 × 205 × 105 mm; optical path length of the samples up to 120 mm
Control and indicationbuilt-in computer: (LAN (2 slots), USB (2 slots), RS 232, VGA, PS/2 keyboard, PS/2 “mouse” manipulator);
built-in display: color, VGA, TFT, LCD, 6,4 inch
Dimensionsnot more than (length × depth × height) 520 × 360 × 210 mm
Weightnot more than 18 kgnot more than 19 kgnot more than 17 kg
Power requirementsfrom 90 up to 255 W, 50-60 Hz, 100 VAfrom 90 up to 255 W, 50-60 Hz, 70 VA

Model МС 122

  • Operating spectral range: from 190 up to 1100 nm
  • Spectral slit width: 3 nm
  • Optical scheme type: monochromator with double dispersion, double-beam (split beam)
  • Light sources: deuterium and halogen lamps
  • Photodetectors: silicon photodiodes
  • Photometric range: by optical density: from -0,3 up to 4 В; by transmittance: from 0,01 up to 200 %
  • Wavelength setting error: not more than ± 0,5 nm
  • Wavelength setting reproducibility: not more than 0,2 nm
  • Wavelength calibration: automatic
  • Margin of error measuring transmittance T: ± 0,3 %
  • Standard deviation measuring transmission T: not more than 0,05 % (Т = from 1 up to 100 %)
  • Zero drift (500 nm): not more than ± 0,0005 В/hour
  • Interfering radiation level: not more than 0,03 % (220, 340, 400 nm)
  • Cuvette compartment: inside dimensions 120 × 205 × 105 mm; optical path length of the samples up to 120 mm
  • Control and indication: built-in computer: (LAN (2 slots), USB (2 slots), RS 232, VGA, PS/2 keyboard, PS/2 “mouse” manipulator); built-in display: color, VGA, TFT, LCD, 6,4 inch
  • Dimensions: not more than (length × depth × height) 520 × 360 × 210 mm
  • Weight: not more than 18 kg
  • Power requirements: from 90 up to 255 W, 50-60 Hz, 100 VA

Model МС 122М

  • Operating spectral range: from 190 up to 1100 nm
  • Spectral slit width: 1,4 nm
  • Resolution for toluene in hexane: greater than 1,7
  • Optical scheme type: monochromator with double dispersion, double-beam (split beam)
  • Light sources: deuterium and halogen lamps
  • Photodetectors: silicon photodiodes
  • Photometric range: by optical density: from -0,3 up to 4 В; by transmittance: from 0,01 up to 200 %
  • Wavelength setting error: not more than ± 0,3 nm
  • Wavelength setting reproducibility: not more than 0,1 nm
  • Wavelength calibration: automatic
  • Margin of error measuring transmittance T: ± 0,3 %
  • Standard deviation measuring transmission T: not more than 0,1 % (Т = from 1 up to 100 %)
  • Zero drift (500 nm): not more than ± 0,0005 В/hour
  • Interfering radiation level: not more than 0,02 % (220, 340, 400 nm); not more than 0,5 % (200 nm)
  • Cuvette compartment: inside dimensions 120 × 205 × 105 mm; optical path length of the samples up to 120 mm
  • Control and indication: built-in computer: (LAN (2 slots), USB (2 slots), RS 232, VGA, PS/2 keyboard, PS/2 “mouse” manipulator); built-in display: color, VGA, TFT, LCD, 6,4 inch
  • Dimensions: not more than (length × depth × height) 520 × 360 × 210 mm
  • Weight: not more than 18 kg
  • Power requirements: from 90 up to 255 W, 50-60 Hz, 100 VA

Model МС 122А

  • Operating spectral range: from 190 up to 1100 nm
  • Spectral slit width: from 0,3 up to 3,0 nm (step 0,1 nm)
  • Resolution for toluene in hexane: greater than 1,9
  • Optical scheme type: monochromator with double dispersion, double-beam (split beam)
  • Light sources: deuterium and halogen lamps
  • Photodetectors: silicon photodiodes
  • Photometric range: by optical density: from -0,3 up to 4 В; by transmittance: from 0,01 up to 200 %
  • Wavelength setting error: not more than ± 0,3 nm
  • Wavelength setting reproducibility: not more than 0,1 nm
  • Wavelength calibration: automatic; by built-in filter
  • Margin of error measuring transmittance T: ± 0,3 % (with a spectral slit width 1,5 nm)
  • Standard deviation measuring transmission T: not more than 0,1 % (Т = from 1 up to 100 %, spectral slit width 1,5 nm)
  • Zero drift (500 nm): not more than ± 0,0005 В/hour
  • Interfering radiation level: not more than 0,02 % (220, 340, 400 nm); not more than 0,5 % (200 nm)
  • Cuvette compartment: inside dimensions 120 × 205 × 105 mm; optical path length of the samples up to 120 mm
  • Control and indication: built-in computer: (LAN (2 slots), USB (2 slots), RS 232, VGA, PS/2 keyboard, PS/2 “mouse” manipulator); built-in display: color, VGA, TFT, LCD, 6,4 inch
  • Dimensions: not more than (length × depth × height) 520 × 360 × 210 mm
  • Weight: not more than 19 kg
  • Power requirements: from 90 up to 255 W, 50-60 Hz, 100 VA

Model МС 124

  • Operating spectral range: from 315 up to 1100 nm
  • Spectral slit width: 3 nm
  • Optical scheme type: monochromator with double dispersion, double-beam (split beam)
  • Light sources: halogen lamp
  • Photodetectors: silicon photodiodes
  • Photometric range: by optical density: from -0,3 up to 4 В; by transmittance: from 0,01 up to 200 %
  • Wavelength setting error: not more than ± 0,5 nm
  • Wavelength setting reproducibility: not more than 0,2 nm
  • Wavelength calibration: automatic
  • Margin of error measuring transmittance T: ± 0,3 %
  • Standard deviation measuring transmission T: not more than 0,1 % (Т = from 1 up to 100 %)
  • Zero drift (500 nm): not more than ± 0,0005 В/hour
  • Interfering radiation level: not more than 0,03 % (340, 400 nm)
  • Cuvette compartment: inside dimensions 120 × 205 × 105 mm; optical path length of the samples up to 120 mm
  • Control and indication: built-in computer: (LAN (2 slots), USB (2 slots), RS 232, VGA, PS/2 keyboard, PS/2 “mouse” manipulator); built-in display: color, VGA, TFT, LCD, 6,4 inch
  • Dimensions: not more than (length × depth × height) 520 × 360 × 210 mm
  • Weight: not more than 17 kg
  • Power requirements: from 90 up to 255 W, 50-60 Hz, 70 VA

Model МС 124-1

  • Operating spectral range: from 380 up to 1100 nm
  • Spectral slit width: 1 nm
  • Optical scheme type: monochromator with double dispersion, double-beam (split beam)
  • Light sources: halogen lamp
  • Photodetectors: silicon photodiodes
  • Photometric range: by optical density: from -0,3 up to 4 В; by transmittance: from 0,01 up to 200 %
  • Wavelength setting error: not more than ± 0,3 nm
  • Wavelength setting reproducibility: not more than 0,1 nm
  • Wavelength calibration: automatic; by built-in filter
  • Margin of error measuring transmittance T: ± 0,3 %
  • Standard deviation measuring transmission T: not more than 0,1 % (Т = from 1 up to 100 %)
  • Zero drift (500 nm): not more than ± 0,0005 В/hour
  • Interfering radiation level: not more than 0,02 % (400 nm)
  • Cuvette compartment: inside dimensions 120 × 205 × 105 mm; optical path length of the samples up to 120 mm
  • Control and indication: built-in computer: (LAN (2 slots), USB (2 slots), RS 232, VGA, PS/2 keyboard, PS/2 “mouse” manipulator); built-in display: color, VGA, TFT, LCD, 6,4 inch
  • Dimensions: not more than (length × depth × height) 520 × 360 × 210 mm
  • Weight: not more than 17 kg
  • Power requirements: from 90 up to 255 W, 50-60 Hz, 70 VA