Device model | МС 122 | МС 122М | МС 122А | МС 124 | МС 124-1 |
---|---|---|---|---|---|
Operating spectral range | from 190 up to 1100 nm | from 315 up to 1100 nm | from 380 up to 1100 nm | ||
Spectral slit width | 3 nm | 1,4 nm | from 0,3 up to 3,0 nm (step 0,1 nm) | 3 nm | 1 nm |
Resolution for toluene in hexane | – | > 1,7 | > 1,9 | – | |
Optical scheme type | monochromator with double dispersion, double-beam (split beam) | ||||
Light sources | deuterium and halogen lamps | halogen lamp | |||
Photodetectors | silicon photodiodes | ||||
Photometric range | by optical density: from -0,3 up to 4 В; by transmittance: from 0,01 up to 200 % | ||||
Wavelength setting error | not more than ± 0,5 nm | not more than ± 0,3 nm | not more than ± 0,3 nm | not more than ± 0,5 nm | not more than ± 0,3 nm |
Wavelength setting reproducibility | not more than 0,2 nm | not more than 0,1 nm | not more than 0,1 nm | not more than 0,2 nm | not more than 0,1 nm |
Wavelength calibration | automatic | automatic | automatic; by built-in filter | automatic | automatic; by built-in filter |
Margin of error measuring transmittance T | ± 0,3 % | ± 0,3 % (with a spectral slit width 1,5 nm) | ± 0,3 % | ||
Standard deviation measuring transmission T | not more than 0,05 % (Т = from 1 up to 100 %) | not more than 0,1 % (Т = from 1 up to 100 %) | not more than 0,1 % (Т = from 1 up to 100 %, spectral slit width 1,5 nm) | not more than 0,1 % (Т = from 1 up to 100 %) | |
Zero drift (500 nm) | not more than ± 0,0005 В/hour | ||||
Interfering radiation level | not more than 0,03 % (220, 340, 400 nm) | not more than 0,02 % (220, 340, 400 nm) not more than 0,5 % (200 nm) | not more than 0,03 % (340, 400 nm) | not more than 0,02 % (400 nm) | |
Cuvette compartment | inside dimensions 120 × 205 × 105 mm; optical path length of the samples up to 120 mm | ||||
Control and indication | built-in computer: (LAN (2 slots), USB (2 slots), RS 232, VGA, PS/2 keyboard, PS/2 “mouse” manipulator); built-in display: color, VGA, TFT, LCD, 6,4 inch | ||||
Dimensions | not more than (length × depth × height) 520 × 360 × 210 mm | ||||
Weight | not more than 18 kg | not more than 19 kg | not more than 17 kg | ||
Power requirements | from 90 up to 255 W, 50-60 Hz, 100 VA | from 90 up to 255 W, 50-60 Hz, 70 VA |
Model МС 122
- Operating spectral range: from 190 up to 1100 nm
- Spectral slit width: 3 nm
- Optical scheme type: monochromator with double dispersion, double-beam (split beam)
- Light sources: deuterium and halogen lamps
- Photodetectors: silicon photodiodes
- Photometric range: by optical density: from -0,3 up to 4 В; by transmittance: from 0,01 up to 200 %
- Wavelength setting error: not more than ± 0,5 nm
- Wavelength setting reproducibility: not more than 0,2 nm
- Wavelength calibration: automatic
- Margin of error measuring transmittance T: ± 0,3 %
- Standard deviation measuring transmission T: not more than 0,05 % (Т = from 1 up to 100 %)
- Zero drift (500 nm): not more than ± 0,0005 В/hour
- Interfering radiation level: not more than 0,03 % (220, 340, 400 nm)
- Cuvette compartment: inside dimensions 120 × 205 × 105 mm; optical path length of the samples up to 120 mm
- Control and indication: built-in computer: (LAN (2 slots), USB (2 slots), RS 232, VGA, PS/2 keyboard, PS/2 “mouse” manipulator); built-in display: color, VGA, TFT, LCD, 6,4 inch
- Dimensions: not more than (length × depth × height) 520 × 360 × 210 mm
- Weight: not more than 18 kg
- Power requirements: from 90 up to 255 W, 50-60 Hz, 100 VA
Model МС 122М
- Operating spectral range: from 190 up to 1100 nm
- Spectral slit width: 1,4 nm
- Resolution for toluene in hexane: greater than 1,7
- Optical scheme type: monochromator with double dispersion, double-beam (split beam)
- Light sources: deuterium and halogen lamps
- Photodetectors: silicon photodiodes
- Photometric range: by optical density: from -0,3 up to 4 В; by transmittance: from 0,01 up to 200 %
- Wavelength setting error: not more than ± 0,3 nm
- Wavelength setting reproducibility: not more than 0,1 nm
- Wavelength calibration: automatic
- Margin of error measuring transmittance T: ± 0,3 %
- Standard deviation measuring transmission T: not more than 0,1 % (Т = from 1 up to 100 %)
- Zero drift (500 nm): not more than ± 0,0005 В/hour
- Interfering radiation level: not more than 0,02 % (220, 340, 400 nm); not more than 0,5 % (200 nm)
- Cuvette compartment: inside dimensions 120 × 205 × 105 mm; optical path length of the samples up to 120 mm
- Control and indication: built-in computer: (LAN (2 slots), USB (2 slots), RS 232, VGA, PS/2 keyboard, PS/2 “mouse” manipulator); built-in display: color, VGA, TFT, LCD, 6,4 inch
- Dimensions: not more than (length × depth × height) 520 × 360 × 210 mm
- Weight: not more than 18 kg
- Power requirements: from 90 up to 255 W, 50-60 Hz, 100 VA
Model МС 122А
- Operating spectral range: from 190 up to 1100 nm
- Spectral slit width: from 0,3 up to 3,0 nm (step 0,1 nm)
- Resolution for toluene in hexane: greater than 1,9
- Optical scheme type: monochromator with double dispersion, double-beam (split beam)
- Light sources: deuterium and halogen lamps
- Photodetectors: silicon photodiodes
- Photometric range: by optical density: from -0,3 up to 4 В; by transmittance: from 0,01 up to 200 %
- Wavelength setting error: not more than ± 0,3 nm
- Wavelength setting reproducibility: not more than 0,1 nm
- Wavelength calibration: automatic; by built-in filter
- Margin of error measuring transmittance T: ± 0,3 % (with a spectral slit width 1,5 nm)
- Standard deviation measuring transmission T: not more than 0,1 % (Т = from 1 up to 100 %, spectral slit width 1,5 nm)
- Zero drift (500 nm): not more than ± 0,0005 В/hour
- Interfering radiation level: not more than 0,02 % (220, 340, 400 nm); not more than 0,5 % (200 nm)
- Cuvette compartment: inside dimensions 120 × 205 × 105 mm; optical path length of the samples up to 120 mm
- Control and indication: built-in computer: (LAN (2 slots), USB (2 slots), RS 232, VGA, PS/2 keyboard, PS/2 “mouse” manipulator); built-in display: color, VGA, TFT, LCD, 6,4 inch
- Dimensions: not more than (length × depth × height) 520 × 360 × 210 mm
- Weight: not more than 19 kg
- Power requirements: from 90 up to 255 W, 50-60 Hz, 100 VA
Model МС 124
- Operating spectral range: from 315 up to 1100 nm
- Spectral slit width: 3 nm
- Optical scheme type: monochromator with double dispersion, double-beam (split beam)
- Light sources: halogen lamp
- Photodetectors: silicon photodiodes
- Photometric range: by optical density: from -0,3 up to 4 В; by transmittance: from 0,01 up to 200 %
- Wavelength setting error: not more than ± 0,5 nm
- Wavelength setting reproducibility: not more than 0,2 nm
- Wavelength calibration: automatic
- Margin of error measuring transmittance T: ± 0,3 %
- Standard deviation measuring transmission T: not more than 0,1 % (Т = from 1 up to 100 %)
- Zero drift (500 nm): not more than ± 0,0005 В/hour
- Interfering radiation level: not more than 0,03 % (340, 400 nm)
- Cuvette compartment: inside dimensions 120 × 205 × 105 mm; optical path length of the samples up to 120 mm
- Control and indication: built-in computer: (LAN (2 slots), USB (2 slots), RS 232, VGA, PS/2 keyboard, PS/2 “mouse” manipulator); built-in display: color, VGA, TFT, LCD, 6,4 inch
- Dimensions: not more than (length × depth × height) 520 × 360 × 210 mm
- Weight: not more than 17 kg
- Power requirements: from 90 up to 255 W, 50-60 Hz, 70 VA
Model МС 124-1
- Operating spectral range: from 380 up to 1100 nm
- Spectral slit width: 1 nm
- Optical scheme type: monochromator with double dispersion, double-beam (split beam)
- Light sources: halogen lamp
- Photodetectors: silicon photodiodes
- Photometric range: by optical density: from -0,3 up to 4 В; by transmittance: from 0,01 up to 200 %
- Wavelength setting error: not more than ± 0,3 nm
- Wavelength setting reproducibility: not more than 0,1 nm
- Wavelength calibration: automatic; by built-in filter
- Margin of error measuring transmittance T: ± 0,3 %
- Standard deviation measuring transmission T: not more than 0,1 % (Т = from 1 up to 100 %)
- Zero drift (500 nm): not more than ± 0,0005 В/hour
- Interfering radiation level: not more than 0,02 % (400 nm)
- Cuvette compartment: inside dimensions 120 × 205 × 105 mm; optical path length of the samples up to 120 mm
- Control and indication: built-in computer: (LAN (2 slots), USB (2 slots), RS 232, VGA, PS/2 keyboard, PS/2 “mouse” manipulator); built-in display: color, VGA, TFT, LCD, 6,4 inch
- Dimensions: not more than (length × depth × height) 520 × 360 × 210 mm
- Weight: not more than 17 kg
- Power requirements: from 90 up to 255 W, 50-60 Hz, 70 VA