MS200
Description
Specification
Configurations
Diffraction gratings

MS200 series monochromator-spectrographs are compact general-purpose spectral devices, completely automated, with enough versatility to satisfy most spectroscopy applications.

We propose several configurations, a wide choice of diffraction gratings, a great number of various accessories and two types of control software.

Choice of configuration:

You can choose necessary device configuration for your task to be solved from the given below list of options or you can consult our specialists on this issue.

Housing of MS200 series monochromator-spectrograph

  • a housing with embedded controller
  • sine mechanism of grating rotation
  • input mirror
  • basic mirrors (focusing and collimator)
  • integrated shutter

Housing options:

  • with standard aluminium (Al+MgF2) coated optics
  • with standard gold coated optics
  • with imaging (astigmatism corrected) aluminium (Al+MgF2) coated optics
  • with imaging (astigmatism corrected device models) gold coated optics
  • with interference optics for all options of the housing, available on request
Housing of monochromator-spectrograph MS200
Housing of monochromator-spectrograph MS200

Diffraction gratings assembly

MS200 series monochromator–spectrographs are available with two types of diffraction gratings mounting:

Automated turret (models MS2004 and MS2004i):

Automated turret in models MS2004 and MS2004i

Up to 4 gratings may be set in the turret. Use of the automated turret allows fast switch of gratings with the help of the control program. Turret axis is orthogonal to the grating rotation axis (wavelength scanning) and provides a high repeatability of the set wavelength while the gratings are switched.

Single grating holder (models MS2001 and MS2001i):

Single grating holder in models MS2001 and MS2001i

It allows easy manual switching of diffraction gratings. A grating is positioned in a holder with high accuracy and no additional tuning of the device is needed while the diffraction grating is changed.

Input port and entrance slit

The devices of this series have one input port on which an entrance spectral slit is installed.

There are two types of slits:

  • Manually operated
  • Automated
All spectral slits have standard coupling flanges providing direct attachment of various accessories of SOL instruments own production to the entrance slit:
  • optical fiber adapter
  • condenser F#-matcher
  • filter wheel
  • fast shutters, etc.
Input port and entrance slit of MS200 monochromator-spectrograph
Input port and entrance slit of MS200

Integrated shutter

All MS200 series monochromator-spectrographs have integrated shutter. It is used for measuring of the background signal or dark signal of photodetector and also for protection of the detectors from high light intensities – the shutter is automatically shut while switching the gratings, filters, zero-order wavelength setting or on change of the output port.

The shutter can be controlled through software or via the external device TTL-signals.

Output ports

MS200 Series monochromators-spectrographs have two output ports – axial and lateral. Installation of an output port is realized with an automated mirror. Either exit spectral slits or adapters for multi-channel detectors can be installed on each output port.

MS200 monochromator-spectrographs can be provided with the following output ports:
  • one axial output port
  • one lateral output port
  • two output ports
Two output ports of monochromator-spectrograph MS200
Two output ports of monochromator-spectrograph MS200

Note:

Due to the small dimensions of the device only detectors with width up to 70 mm may be attached to the axial output port. Detectors intended to be attached to the lateral port have no size limitations.

One axial output port:

One axial output port of MS200 monochromator-spectrograph

This configuration is usually applied if only one detector is used. In this case the lateral port is covered with a cap.

Note
Due to small dimensions of the monochromator-spectrograph only detectors with the width up to 70 mm may be attached to the axial output port.

One lateral output port:

One lateral output port of MS200 monochromator-spectrograph

The lateral output port should be used for the detectors with no opportunity of being attached to the axial output port of the device due to their dimensions. For one lateral output port a fixed turning mirror is used.

In configuration with one lateral output port an axial port is covered with a cap.

Two output ports with mirror:

Two output ports with switching mirror in monochromator-spectrograph MS200

When both output ports are used the automated output port selection mirror is installed in the device for the fast switching between the ports. A high precision of the output mirror positioning (better than 0.03 nm for 1200 l/mm grating) allows obtaining of reproducible spectra at switching the output ports.

Note
Due to small size of the device only detectors with the width up to 70 mm may be attached to the axial output port.

Spectral slits

Monochromator-spectrographs of MS200 series have two types of input and output spectral slits:
  • manually operated
  • automated (combined)
Automated spectral slit of MS200 with attached optical fiber adapter
Automated spectral slit of MS200 with attached optical fiber adapter

Additional equipment:

All spectral slits have standard flanges that make possible attachment of various optical accessories of own production directly to the entrance slit:
  • optical fiber adapter
  • condenser F#-matcher
  • filter wheel
  • fast shutter
  • detectors based on photodiodes and PMT
  • adapter for IR detector
  • other custom accessories

Manually operated spectral slits:

Manually operated spectral slit of monochromator-spectrograph MS200
Widthregulated from 0 mm
up to 2.0 mm
Parallelism± 1 µm
Accuracy
(slit width 1 mm)
± 10 µm
Repeatability± 1.5 µm
Reading accuracy (micrometer)2 µm
Height10 mm
  • Width: regulated from 0 up to 2.0 mm.
  • Parallelism: ± 1 µm.
  • Accuracy (slit width 1 mm): ± 10 µm.
  • Repeatability: ± 1.5 µm.
  • Reading accuracy (micrometer): 2 µm.
  • Height: 10 mm.

The slit width is operated manually by means of the micrometer. Slit height is regulated by a diaphragm in the range from 0 to 10 mm. Diaphragm division value is 0,5 mm.

Automated spectral slits:

Автоматизированная спектральная щель монохроматора-спектрографа MS200
Widthregulated from 0 mm
up to 2.0 mm
Parallelism± 1 µm
Accuracy
(slit width 1 mm)
± 10 µm
Repeatability± 1 µm
Reading accuracy (micrometer)2 µm
Step size0.5 µm
Height10 mm
  • Width: regulated from 0 up to 2.0 mm.
  • Parallelism: ± 1 µm.
  • Accuracy (slit width 1 mm): ± 10 µm.
  • Repeatability: ± 1 µm.
  • Reading accuracy (micrometer): 2 µm.
  • Step size: 0.5 µm.
  • Height: 10 mm.

In the automated spectral slits a stepper motor is used for PC control over the blades opening width. The slit width also may be controlled manually by means of a micrometer screw. The slit height is regulated by a special diaphragm.

Control of the MS200 series monochromator-spectrograph

Monochromator-spectrographs of MS200 series are completely automated spectral devices.

The basic control of the monochromator-spectrographs is executed through the embedded controller. It performs continuing control of commands execution received from the external control device.

Besides, the embedded controller takes control over the built-in ADC which processes signals from the integral photodetectors – PMT, photodiode and etc.

External control of the devices can is performed through PC.

DevCtrl software:

The control software DevCtrl is supplied with the monochromator-spectrograph.

DevCtrl controls over all drivers and units of the device and accessories. The software also gives an access to the system parameters and setup coefficients of the device.

DevCtrl - control software for MS200 monochromator-spectrograph

Spectroscopy software SpectraSP:

Spectroscopy software SpectraSP is intended for joint control over monochromator-spectrographs and detection systems based on the integral or multi-element photodetectors. SpectraSP provides simultaneous control over the device and optical radiation detection process by means of various detectors with subsequent processing of acquired data and their graphical representation.

More about SpectraSP

Via SpectraSP is performed:

  • control of the device units and mechanisms
  • registration of optical radiation using various detectors
  • processing of spectral measurements results
  • visualization of the acquired results in the form of a graphical image of the spectra
Spectroscopy software SpectraSP

Diagrams of possible device configurations

Accessories connected to the input port of MS200
Accessories connected to the input port of MS200
Accessories connected to the output port of MS200
Accessories connected to the output port of MS200

Additional equipment: